Abstract
AbstractAn overview of the existing methods for minimization of the analysis time in gas chromatography (GC) is presented and a new system for fast temperature programming and very fast cooling down is evaluated. In this study, a system of coaxial tubes, a heating/cooling module (HC-M), was developed and studied with a capillary column placed inside the HC-M. The module itself was heated by a GC oven and cooled down by an external cooling medium. The HC-M was heated at rates of up to 330 °C min−1 and cooled at the rate of 6000 °C min−1. The GC system was prepared for the next run within a few seconds. The HC-M permits good separation reproducibility, comparable with that of a conventional GC, expressed in terms of relative retention times and peak areas of analytes reproducibilities. The HC-M can be used within any commercial gas chromatograph.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.