Abstract

The gas Chromatographic behaviour of the ternary complexes of selected bivalent first-row transition metal ions with 1,1,1,5,5,5-hexafluoro-2,4-pentanedione H(HFA), and di-n-butylsulphoxide, DBSO, was studied. Calibration plots of peak area vs. amount of metal injected were linear over a range of 60-900 ng for manganese(II), iron(II), cobalt(II) and nickel(II). The average relative standard deviation was less than 3.0% for all the metals studied. Detection limits of 60, 109, 112 and 115ng for cobalt(II), nickel(II), iron(II) and manganese(II), respectively, were obtained with flame-ionization detection. Various liquid phases, including OV-1, SE-30, and Dexsil 300 were used. The best results were obtained on columns of 5% Dexsil 300. No appreciable thermal decomposition was observed on stainless-steel or glass columns, but the best formed peaks were obtained on all-glass columns. The elution of the metallic species was confirmed by venting the exit gases from the gas chromatograph directly into an atomic-absorption spectrophotometer.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.