Abstract
The gas chromatographic–mass spectrometric (GC–MS) separation of all 209 polychlorinated biphenyl (PCB) congeners was studied on an extremely efficient 80 m × 0.1 mm i.d. capillary column coated with a 0.1 μm film of poly(5%-phenyl methyl)siloxane stationary phase. The quality of the separation and the number of resolved and coeluting peaks were compared to predictions according to the statistical overlap theory (SOT) and to literature data on PCB separations obtained by one-dimensional and comprehensive two-dimensional GC (GC × GC) and GC–MS. Mass spectral and chemometric deconvolution procedures were used to resolve overlapping peaks. On the highly efficient column, 195 PCB congeners were resolved in 96 min separation time using spectral and chemometric deconvolution. This number is comparable to the best separations described in GC × GC–MS mode. The novel method was developed for spectral deconvolution of overlapped PCB congeners which was verified determining the most toxic, dioxin-like PCBs both in the model mixture of 209 PCBs as well as in the Aroclor 1242 and Aroclor 1254 formulations.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.