Abstract

A sensitive and specific method for the quantitative determination of morphine in human plasma is presented. Morphine was extracted from plasma by solid phase extraction on C18 and converted to its pentafluorobenzyl carbonate trimethylsilyl derivative. The derivatives were analysed without further purification. Using gas chromatography/negative ion chemical ionisation mass spectrometry, a useful diagnostic fragment ion at m/z 356 is obtained at high relative abundance. Deuterated morphine was used as internal standard. Calibration graphs were linear within the range 1.25 to 320 nmol/L. Intra-day precision was 3.82% (15 nmol/L), 2.85% (75 nmol/L) and 4.13% (225 nmol/L), inter-day variability was found to be 1.77% (15 nmol/L), 4.95% (75 nmol/L) and 9.88% (225 nmol/L). Inter-day accuracy showed deviations of 2.18% (15 nmol/L), -0.72% (75 nmol/L) and -0.13% (225 nmol/L). The method is rugged and robust and has been applied to the batch analysis of morphine during pharmacokinetic profiling of the drug.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.