Abstract
A GC–HRMS based method for the accurate and sensitive determination of nine organotin compounds, tetrabutyltin (TeBT), tributyltin (TBT), dibutyltin (DBT), monobutyltin (MBT), triphenyltin (TPhT), diphenyltin (DPhT), monophenyltin (MPhT), tricyclohexyltin (TCyT), and dicyclohexyltin (DCyT) in sediment, tissue and water samples is presented and discussed. Mass spectral features of these analytes via both low resolution quadrupole and high resolution magnetic sector, GC–HRMS conditions under selective ion monitoring mode and QA/QC criteria for the positive identification of analyte are all provided. Linearity of response and minimal detectable limits are illustrated for each of the nine compounds monitored and the estimates of method limits-of-detection were 7–29 ppt for water and 0.35–1.45 ppb for tissue or sediments. Sample preparation considerations and precision are discussed for spiked water and sediment samples, whereas method accuracy was established by analysing a certified reference material (CRM) mussel sample and comparing our results to the assigned values. Good agreement was found between our results and assigned or indicative values for MBT, DBT, TBT, DPhT and TPhT (cyclohexyl-tins were not present in the CRM).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.