Abstract

A very sensitive method for the determination of acetone, diacetyl, and CᾢC alcohols by gas chromatography has been described. With 30% (w/w) polyethylene glycol 400 on celite 545 (60/80 mesh) support and a column temperature of 70°C, 0.5 μg of acetone, methanol, ethanol, iso-propanol, 0.25 μg of n-propanol, iso-butanol, n-butanol and 1.2 μg diacetyl were detected on the hydrogen flame detector. At a column temperature of 110°C, 0.1 μg of n-pentanol, n-hexanol and n-heptanol were the minimum amounts detected. The variation and reproducibility was 3% at all concentrations tested (up to 300 or 500 μg). Peak height measurements were found superior to peak area measurements.