Abstract
A method is described for the determination of pentachlorophenol (PCP) in gelatin. The method employs acid and heat to hydrolyze the gelatin matrix, a base partition and wash for separation and cleanup, and a reacidification and extraction with hexane for direct determination of PCP, without preparation of a derivative, using gas chromatography (GC) with a 1% SP- 124ODA liquid phase and a 63Ni electron capture detector. Recoveries averaged 106% for fortifications between 0.02 and 1.0 ppm. The limit of quantitation is 20 ppb. The limit of detection is 4-6 ppb. The method, which has undergone a successful intralaboratory trial, is simple and rapid, and requires only general laboratory reagents and equipment. GC of the acetate derivative of PCP is used for confirmation of identity.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.