Abstract

Ten laboratories participated in a collaborative study of a method for the determination of deoxynivalenol in wheat by gas chromatography with electron capture detection. Each laboratory analyzed 6 samples in duplicate. Each collaborator received samples spiked at the 100.3, 501.3, and 1002.6 ng/g levels; a control sample; and 2 naturally contaminated samples. The average recovery (outliers excluded) for the spiked samples was 92.2%. The mean repeatability and reproducibility, respectively, were 32.2 and 41.3% for the spiked samples and 30.9 and 47.6% for the naturally contaminated samples. The method was adopted official first action.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.