Abstract

Gas chromatographic determination of azinphosmethyl and its thermally labile metabolite, azinphosmethyl-oxon, can be achieved with cool on-column injection and element-specific flame photometric detection. This paper presents a method of analysis for airborne residues of azinphosmethyl and azinphosmethyl-oxon in XAD-4 resin extracts using cool on-column injection. Calculated limits of quantitation for the analytical method are 8.1 ng/m3 for azinphosmethyl and 10.2 ng/m3 for azinphosmethyl-oxon. Average percentage recovery and percentage relative standard deviations for 10 replicates were 93±18% for azinphosmethyl and 104±16% for the oxon. This analytical method is rapid, robust and requires a minimum of gas chromatograph injection port maintenance.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.