Abstract

The performance of a fused silica column coated with 50% phenyl-/50% dimethyl-polysiloxane, characterized by a high thermal stability, was evaluated for the gas chromatographic determination of cholesterol oxidation products. A silylated mixture of eighteen oxysterol standards (0.1 mg/ml of each compound) was injected into the gas chromatographic system. The temperature was programmed from 230 to 260°C at 2.5°/min and then to 290°C at 1°/min; the injector and detector temperatures were set at 325°C, and the gas linear velocity was 22.8 cm/s. The column gave a fast (15 min) and satisfactory resolution of the main cholesterol oxides. Good separation of the hydroxy from the epoxy derivatives was achieved.The suitability of the method for the determination of oxysterols in food matrices was successfully tested on a saponified lipid extract from egg yolk powder, previously enriched and purified by NH2 solid phase extraction.