Measurement of linear polarization of gamma-rays from nuclei oriented by hyperfine interaction at low temperatures using the3He/4He dilution refrigerator at the SERC Daresbury Laboratory is reported. Three different types of a Compton polarimeter were assembled using coaxial Ge(Li) detectors. The method was applied to192IrFe and60CoFe and a polarization detection efficiency curve and a figure of merit for each polarimeter were produced. As a result of combined directional distribution and linear polarization measurements multipole mixing ratios of several transitions in192Pt have been obtained and are compared with theoretical predictions.