Abstract

The appearance of small-sized and powerful enough electric cryocoolers of various types on the market, has opened the perspective of HPGe detectors application, cooled by such coolers, in radiation portal monitors. The first results of a spectrometric module based on HPGe detector with relative efficiency of 45% cooled by a Stirling-cycle cryocooler, are presented. The spectrometer has provided energy resolutions of less than 0.95 keV and 1.95 keV at energies of 122 keV and 1332 keV, respectively. The deterioration of the energy resolution of HPGe detector cooled by electric cryocooler in comparison to the resolution with liquid nitrogen cooling was about 8% at the energy of 1332 keV. With the use of activated filters to suppress pulses produced by the mechanical vibrations, the energy resolution of the spectrometer was 0.8 keV and 1.8 keV, respectively, however, the detector relative efficiency at the energy of 1332 keV has dropped to 39 %.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.