Abstract

In order to clarify the γ-efficiency dependency of 4πβ-γ efficiency functions, a series of 4πβ-γ efficiency extrapolation measurements of a 134Cs source were carried out for a wide variety of γ-geometries using a 4πβ(PS)-4πγ detector configuration. As the source is situated in the plastic scintillator (PS) β-detector, the γ-efficiency of the system can be readily changed by extracting the β-detector from the well-hole in a series of stages. For data acquisition and analyses, a list-mode two-parameter data acquisition system was employed. The forms of the extrapolation curves were monitored with decreasing γ-geometry, eventually exhibiting a similar behavior to those obtained in a usual 4πβ-γ coincidence counting system. The experimental results and considerations suggested that the γ-geometry dependency of the efficiency functions were due to summing effects in the γ-channel, and some qualitative remarks on the form of the extrapolation functions are given.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.