Abstract
We developed a positron intraoperative probe with high sensitivity dedicated to real-time assessment of the resection margins of tumors and control of the surgical cavity. Its main limitation comes from the annihilation photon background generated by distal non-specific radiotracer uptakes. This study is a proof of concept for a dual-scintillator intraoperative probe with background subtraction capability. It associates a highly p-terphenyl β-sensitive scintillator and a highly γ-sensitive LYSO:Ce. Both scintillator are readout by a single pixelated photodetector based on SiPMs and the shape and intensity of the light distribution is used to identify in which scintillator the interactions occurred to allow further discrimination of the signals and subsequent quantification of this background can be utilized to have it subtracted from the raw images. We describe in this study the design of the probe and the analysis to achieve this identification with a good accuracy. An average efficiency of events identification of 97.90 ± 0.06% is reported on the whole field of view.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.