Abstract
In hemiparetic patients with an implantable peroneal stimulator for correction of drop foot the gait pattern was studied over several years. The gait parameters and M-waves of subcutaneously stimulated muscles were compared with the results obtained before implantation and their variation was observed over time. Of a group of 35 patients with previously implanted electrodes 19 were evaluated. Significant improvements of gait were found although in some cases an excessive eversion of the foot was observed. Nine of these patients had reimplantation because of displacement of the stimulation electrodes after an average time of 3.5 years of proper functioning of the implant. After the reimplantations, similar gait patterns and muscular responses to stimulation were observed as after the initial implantation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.