Abstract

A fuzzy logic-based state of health (SOH) meter is being developed for lithium-ion (Li-ion) batteries for potential use in portable defibrillators. Electrochemical impedance spectroscopy (EIS) measurements have been made from which input parameters for a fuzzy logic model to estimate the state of charge (SOC) and SOH are derived. The batteries are discharged continuously at a 1.4 A load current to simulate the constant current draw during the monitoring and recording of a patient's EKG, and periodically interrupted by 10 A pulses to simulate the battery discharge to charge up the capacitor that is in turn discharged to supply high voltage to the electrodes for the defibrillation of the patient. The test procedures included both voltage recovery and EIS measurements, and were made as the batteries were being discharged and over 30 charge/discharge cycles. Accurate models have been developed to estimate the number of pulses that the battery pack can deliver at various stages of its cycle life (SOC measure) and the number of charge/discharge cycles (SOH measure) that it had undergone.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call