Abstract

In this study, using the fuzzy logic method, a stress detection tool was created with body temperature and blood pressure parameters as indicators to determine a person's stress level. This tool uses the LM35DZ sensor to detect body temperature, the MPX5100GP sensor to read blood pressure values, and Arduino Uno as a data processor from sensor readings which are then calculated using the fuzzy logic method as a stress level decision-maker. The resulting output measures blood pressure, body temperature, and the stress level experienced by a person, which will be displayed on the liquid crystal display. Based on the results of testing the body temperature parameter, the highest error generated was 1.17%, and for the blood pressure parameter, the highest error was 2.5% for systole and 0.93% for diastole. Furthermore, testing the stress level displayed on the tool is compared to the depression, anxiety, and stress scales 42 (DASS 42), a psychological stress measuring instrument. From the results of testing the tool with the questionnaire, the average conformity level is 74%.