Abstract

This work builds upon previous investigations of lung tumour peripheral doses for 6 MV, 6 MV FFF, and 10 MV FFF conformal arc therapy beams calculated in the Monaco® TPS and delivered using an Elekta® Agility™ linac. An improved patient lung phantom is developed with measurements using the normoxic PAG gel dosimeter and compared against dose planes from the TPS. The gel dosimeter measurements indicate that the TPS is overestimating the secondary build-up in the lung tumour peripheral region. It has been determined that in lung tumours, 6 MV FFF is the optimal beam energy for peripheral dose coverage and that there is a dosimetric compromise using 10 MV FFF.