The Sussex Mk4 Electrical Impedance Mammography (EIM) system is a novel instrument, designed for the detection of early breast cancer, based upon Electrical Impedance Tomography (EIT). Many innovations in the field have been incorporated in the design improving both signal distribution and response. This paper investigates the behaviour of the contactless patient-electrode interface. The interface was studied in detail using phantom and healthy volunteer, in-vivo, data. Our findings show the necessity for the careful design of electrode enclosure so that the response of the system is not affected by the unpredictable positioning of the breast; it closely mimics those conditions seen when using the phantom. The paper includes a number of possible designs and their individual characteristics. In addition an explanation on the unanticipated effects and solutions for such are described.