Abstract

Analysis of calibration data of a rotary-vane attenuator has yielded closer agreement between the recently developed off-null and the proven power ratio methods of attenuation measurement. A constant bias was discovered in the measured valures of attenuation difference, and a procedure is described to correct for this bias. After the correction is applied, the average agreement between the two methods of measurements is improved an order of magnitude, namely, from a former 2.0 percent to 0.17 percent.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.