Abstract
An attempt is made at an estimate of the ultimate limitations to the precision of mass comparison by means of a beam balance. Systematic effects due to imperfections of the balance suspensions, air buoyancy, humidity and load decentring are not considered. Mathematical models of knife-edge balances and flexure-strip balances are used to derive balance sensitivity and susceptibilities to tilt and horizontal ground vibration. It is shown that optimum designs exist and the magnitude of ground vibration perturbation is calculated. A rigorous treatment of thermal noise is given and the magnitude of the fluctuating torques due to viscous damping is calculated. The principal conclusion is that the precision of weighing is at present probably limited by ground vibration noise. Nevertheless, knife-edge balances should be capable of a resolution of tens of nanograms.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Proceedings of the Royal Society of London. A. Mathematical and Physical Sciences
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.