Abstract

The signal-to-noise ratio of a contactless conductivity detector for capillary electrophoresis was examined for different cell arrangements and operating parameters. The best signal-to-noise ratios, and hence the best detection limits, are obtained for frequencies which give highest sensitivity. Comparative experiments for three different excitation voltages (20, 100, and 200 V(pp)) showed that the best signal-to-noise ratios were achieved for the highest excitation voltage of 200 V(pp). Low conductivity of the background electrolyte solution is mandatory to obtain lowest noise levels, and also the improvement on applying high excitation voltages was best for the electrolyte solution with lowest conductivity. The diameter of the electrodes was found to have only a negligible effect, so that a tight fitting of the electrodes to the external diameter of the capillary is not necessary. A cell without shielding between the two electrodes showed significant direct coupling (stray capacitance) and lower signal-to-noise ratios for all experimental conditions used. A serious distortion of the peak shapes was also observed for this cell arrangement.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.