Abstract

BackgroundDesign freedom offered by additive manufacturing allows for the implementation of functional gradients - where mechanical stiffness is decreased along the length of the implant. It is unclear if such changes will influence failure mechanisms in the context of rib fracture repair. We hypothesized that our novel functionally graded rib implants would be less stiff than controls and decrease occurrence of secondary fracture at implant ends. MethodsFive novel additively manufactured rib implants were tested along with a clinically used Control implant. Fracture reconstructions were modeled with custom synthetic rib bones with a transverse B1 fracture. Ribs were compressed in a cyclic two-point bend test for 360,000 cycles followed by a ramp to failure test. Differences in cyclic stiffness, 3D interfragmentary motions, ramp-to-failure stiffness, maximum load, and work to failure were determined. FindingsThe Control group had lower construct stiffness (0.76 ± 0.28 N/mm), compared to all novel implant designs (means: 1.35–1.61 N/mm, p < 0.05) and rotated significantly more about the bending axis (2.7° ± 1.3°) than the additively manufactured groups (means between 1.2° – 1.6°, p < 0.05). All constructs failed via bone fracture at the most posterior screw hole. Experimental implants were stiffer than Controls, and there were few significant differences between functional gradient groups. InterpretationAdditively manufactured, functionally graded designs have the potential to change the form and function of trauma implants. Here, the impact of functional gradients was limited because implants had small cross-sectional areas.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call