Abstract

Contrast between blood and myocardium in standard turbo gradient echo MR techniques (TFE) used routinely in clinical practice is mainly caused by unsaturated inflowing blood. Steady-state free precession (SSFP) has excellent contrast even in the absence of inflow effects. In 45 subjects cardiac cine loops in two long axis projections were acquired using TFE and compared with SSFP. A visual score (range 0 worst - 3 best) was assigned for endocardial border delineation for six myocardial segments in two long axis views. Endocardial border delineation score for TFE was 1.3 +/- 0.3 per segment and 2.4 +/- 0.3 for SSFP (P < 0.0001). Signal intensity blood/signal intensity myocardium was 1.5 +/- 0.4 at enddiastole and 1.4 +/- 0.3 at systole for TFE and 3.5 +/- 1.1 and 3.2 +/- 1.3 for SSFP, respectively (P < 0.0001). SSFP increases contrast between blood and myocardium more than twofold, resulting in an improved endocardial border definition. This may reduce variability for the determination of cardiac volumes and ejection fraction.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.