Abstract

Implantable systems are often employed to perform continuous high-resolution recordings of neural activity. These systems frequently require invasive procedures when implanting and maintaining effective operation. This causes major interruptions to daily life. Previous work demonstrated an in vitro minimum detectable signal of 15 μ V in amplitude and RF sensitivity down to –135 dBm. This suggests the possibility of detecting diminutive biopotentials in a wireless fully passive manner. Here, for the first time, we validate the wireless neurosensing system through a series of in vivo electrophysiological recordings including both spontaneous cardiac activity and sensory evoked neural activity, with amplitudes ranging from a few microvolts to millivolts and across a spectrum of frequencies. We also present design considerations and the development of probes for neurosensing to accomplish detectability of biopotentials in the tens of microvolts in rats. The developed probes show improved impedance matching with the neurosensing system. Specifically, the new probes showed an impedance several orders of magnitude lower than those commercially available, thereby significantly improving signal detection. Notably, the presented in vivo validation of this technology has great future clinical implications in neuroscience as it offers a wireless and unobtrusive device for neurological research, monitoring, and therapeutic purposes.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call