Abstract
Microfluidic chip analysis has great potential advantages such as high integration, fast speed analysis, and automatic operation and is widely used not only in biological fields but also in many other analytical areas such as agriculture and food safety. Herein, a fully automatic multi-class multi-residue analysis of veterinary drugs simultaneously in an integrated chip-mass spectrometry (chip-MS) platform was developed. The developed microfluidic chip platform integrated three modules including the extraction and filtration module, "pass-through" clean-up module, and online evaporation module. The resulting chip has been coupled to a MS detector successfully, in which 23 kinds of residues in five classes were simultaneously qualitatively and quantitatively detected without chromatographic separation, obtaining the limits of detection of the spiked milk sample in the range of 0.23-4.13 ng/mL and the recovery rate in the range from 71.7 to 118.0% under optimized conditions. The microfluidic chip system developed in this study provided a new idea for the development of detection chips and exhibited considerable potential in the point-of-care testing in milk.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.