Abstract

For routine clinical implementation of Alzheimer's disease (AD) plasma biomarkers, fully automated random-access platforms are crucial to ensure reproducible measurements. We aimed to perform ananalytical validation and to establishcutoffs for AD plasma biomarkers measured with Lumipulse. Two cohorts were included.UNIPG: n=450 paired cerebrospinal fluid (CSF)/plasma samples from subjects along the AD-continuum, subjects affected by other neurodegenerative diseases, and controls with known CSF profile;AMS: n=40 plasma samples from AD and n=40 controls. Plasmaamyloid β (Aβ)42, Aβ40, and p-tau181 were measured with Lumipulse. We evaluated analytical and diagnostic performance. Lumipulse assays showed high analytical performance. Plasma p-tau181 levels accurately reflected CSF A+/T+ profile in AD-dementia and mild cognitive impairment (MCI)-AD, but not in asymptomatic-AD. Plasma and CSF Aβ42/40 values were concordant across clinical AD stages. Cutoffs and probability-based models performed satisfactorily in both cohorts. The identifiedcutoffs and probability-based models represent a significant step toward plasma AD moleculardiagnosis.

Full Text
Published version (Free)