Abstract

Herein, we describe the development of a fully automatable technology that features online coupling of high-pH RP separation with conventional low-pH RP separation in a two-dimensional capillary liquid chromatography (2-D LC) system for shotgun proteomics analyses. The complete analysis comprises 13 separation cycles, each involving transfer of the eluate from the first-dimension, high-pH RP separation onto the second RP dimension for further separation. The solvent strength increases across the 13 fractions (cycles) to elute all peptides for further resolution on the second-dimension, low-pH RP separation, each under identical gradient-elution conditions. The total run time per analysis is 52 h. In triplicate analyses of a lysate of mouse embryonic fibroblasts, we used this technology to identify 2431 non-redundant proteins, of which 50% were observed in all three replicates. A comparison of RP-RP 2-D LC and strong cation exchange-RP 2-D LC analyses reveals that the two technologies identify primarily different peptides, thereby underscoring the differences in their separation chemistries.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.