Because HIV medications are used in combination, it is important to develop multiplex assays to streamline the therapeutic drug monitoring process and provide rapid turnaround. This article reports full validation of an analytical method that combines atazanavir with 6 HIV-protease inhibitors (indinavir, amprenavir, saquinavir, nelfinavir, ritonavir, and lopinavir) and 2 nonnucleoside reverse transcriptase inhibitors (nevirapine and efavirenz). Using 200 microL of plasma and a simple liquid-liquid extraction method, this analytical method achieved a clean baseline and high extraction efficiencies (90.0% to 99.5%). A Zorbax C-18 (150 x 4.6 mm, 3.5 microm) analytical column was used along with a 27-minute linear gradient elution of the mobile phase to provide sharp peaks at 210 nm. This method was validated over a range of 25 to 10,000 ng/mL and is accurate (90.4% to 110.5%) and precise (precision within a day and between days ranged from 2.3% to 8.3%). Because this method is simple and inexpensive, it may have applicability in countries with low resources.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.