Abstract

Purpose: To validate a bioanalytical inductively coupled plasma mass spectrometry (ICP-MS) method for quantification of iron in human plasma with K2EDTA (matrix). This method was validated using typical bioanalytical validation parameters. Materials and methods: Iron reference standards and germanium internal standard (IS) were from Ultra Scientific. Reagent water had a resistivity ?18 M?cm. Nitric acid and hydrochloric acid were ultra pure grade. Blank matrix was purchased from BioReclamation. Linearity was assessed using eight standards prepared in proxy matrix over a range equivalent to 0.5-20 ?g iron/mL of matrix at the beginning and end of each run. Quality control (QC) samples were prepared by fortifying matrix at 0.5, 1.5, 10, 15 and 100 ?g/mL. A blank containing IS and a double blank were included in every run. Selectivity was evaluated using six different lots of matrix fortified with IS and analyte at the LLOQ level. A Thermo-Finnigan ELEMENT2 High Resolution ICP-MS was used to analyze prepared samples. Results: System suitability, selectivity, carryover, and ruggedness all met the acceptance criteria. Linearity was consistently demonstrated. Recoveries for all calibration standards and QCs were within the acceptance criteria. Freeze/thaw was demonstrated for 4 cycles. Bench top reproducibility was successful following 53 hours of ambient storage. Reinjection reproducibility was successful following 96 hours of ambient storage. Extract stability was successful following 190 hours of ambient storage. IS solution stability was successful following 29 days of ambient storage. Calibration standard stability was successful following 8 days of ambient storage. Analyte stock solution stability was successful following 29 days of ambient storage. Long term storage stability was successful following 31 days at 2 to 8°C, 31 days at -15 to -25°C, and 30 days at ?70 to -90°C storage. Conclusion: This bioanalytical ICP-MS method was successfully validated for the quantification of iron in human plasma with K2EDTA.

Highlights

  • Some heavy metals represent a serious health risk and are eliminated very slowly from the body

  • Long term storage stability was successful following 31 days at 2 to 8°C, 31 days at -15 to -25°C, and 30 days at ‐70 to -90°C storage. This bioanalytical inductively coupled plasma mass spectrometry (ICP-MS) method was successfully validated for the quantification of iron in human plasma with K2EDTA

  • An high resolution (HR) inductively coupled plasma (ICP)-MS assay for iron in human plasma with K2EDTA was successfully validated by MPI Research

Read more

Summary

Introduction

Some heavy metals represent a serious health risk and are eliminated very slowly from the body. Heavy metal analysis has been performed under the auspices of the Environmental Protection. Analytical techniques were based primarily on qualitative, non-specific, low-sensitivity colorimetric procedures based on the precipitation of insoluble metal sulfides. Inductively coupled plasma mass spectrometry (ICP-MS) has been used to provide accurate, sensitive, specific quantification of heavy metals in environmental samples, forensic samples, nutraceutical products, and drug articles (drug substances, drug products, and excipients)[4,5,6,7,8,9]. The main advantages of ICP-MS over the classical methods are its ability to accurately quantify trace metals at much lower detection limits, analytical speed, relative lack of interference, and definitive multiple isotope capability. The USP has identified four heavy metal contaminants known or suspected to pose a significant

Objectives
Methods
Results
Conclusion