Abstract
Interferometry associated with an external cavity laser of long coherence length and broad wavelength tuning range shows promising features for use in measurement of absolute distance. As far as we know, the processing of the interferometric signals has until now been performed by Fourier analysis or fringe counting. Here we report on the use of an autoregressive model to determine fringe pattern frequencies. This concept was applied to an interferometric device fed by a continuously tunable external-cavity laser diode operating at a central wavelength near 1.5 microm. A standard uncertainty of 4 x 10(-5) without averaging at a distance of 4.7 m was obtained.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.