Abstract

Short-duration, cyclically loaded, axial, fretting corrosion tests were designed and performed to compare the fretting behaviour of different metal Howmedica® universal modular heads connected to coated and uncoated metal cone taper spigots. Concurrent axial push-on and pull-off taper friction tests were also performed. There was no measurable fretting debris found in any test and SEM studies of the contact surfaces showed no evidence of fretting. It is concluded that no significant, long-term fretting corrosion of the Vitallium® Co-Cr-Mo heads on the Vitallium or uncoated Ti-6AI-4V spigots, nor of the Orthinox® stainless steel heads on the Orthinox spigots, is likely to occur in vivo.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.