Abstract

A cw dye laser system frequency stabilized to a high-finesse optical reference cavity is described. Laser frequency is servo controlled to the cavity resonance with residual fluctuations less than 50 kHz for short times (20 μsec) and 100 Hz for long times (10 sec). Drift in absolute laser frequency of about 1.5 MHz/min is observed due to drift of the unstabilized reference cavity. A saturated absorption spectrum of I2 obtained with this system is shown.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.