Piezoelectric sensors are used to pick up low level vibrations in automotive, biomedical and industrial fields and are subject to errors due to the input current and voltage noise of amplifiers, when connected with a large bias resistor. The noise levels, especially from the bias resistor, degrade the signal to noise ratio of the signal which leads to less reliable feature detection. We report the experimental investigation of a frequency sensing front-end in terms of its linearity, noise and gain in order to acquire medical signals at low frequencies and voltages, e.g. arterial pulse signal, Ballistocardiogram and heart sounds. An LC oscillator is used with a commercial phase locked loop for this purpose. We show that the noise levels in frequency sensing can be reduced by increasing the oscillation frequency, while maintaining 1% non-linearity. Overall, the readout runs at 5V and has a mean noise floor of 1.43μV/Hz and a non-linearity of 1% when subject to a sine wave between 20mV and 200mV peak to peak amplitude. The results of this work are expected to contribute towards low noise analog front end designs for piezoelectric sensors using frequency sensing as an alternate architecture.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.