Abstract

The resonant frequencies of five commonly used fluid-filled catheters connected to a solidstate microdisplacement pressure gauge were 18 to 33 per cent higher than those obtained with a conventional manometer. Four catheters had a flat amplitude response to 26 Hz. or more at the 10 per cent amplitude distortion level. The dynamic response characteristic of certain fluid-filled catheters used with microdisplacement pressure gauges allows one to record high fidelity left ventricular pressure curves.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.