Abstract

The optically pumped cesium beam clock named Cs IV is operated with a new short Ramsey cavity satisfying strict requirements on the microwave leakage level. The most relevant characteristics of the device are presented. Cs IV is presently driven by standard electronics coming from a HP 5061 B clock that provides a sinusoidal modulation of the interrogation microwave signal and a microwave power stability of about 1% at a temperature of 20+/-1 degrees C. The short- and medium-term frequency stability measurement gives sigma(y)(1 day)=2x10(-14): this value holds up to 3 days. The accuracy evaluation results in an uncertainty of 10(-12), and the repeatability is evaluated to 3x10(-13). It appears that the flicker floor is beginning at 2x10(-14) and is mainly due to both the power fluctuations of the free running microwave interrogating signal and the fluctuations of the external static magnetic field. The accuracy is limited by the lack of knowledge of the end-to-end cavity phase shift.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.