Abstract

Our study proposes the use of a frequency-modulated acquisition which suppresses banding artefacts in combination with a phase-sensitive water-fat separation algorithm.The performance of the phase-sensitive separation for standard bSSFP, complex sum combination thereof, and frequency-modulated bSSFP were compared in in vivo measurements of the upper and lower legs at 1.5 and 3 T. It is shown, that the standard acquisition suffered from banding artefacts and major swaps between tissues. The dual-acquisition bSSFP could alleviate banding artefacts and only minor swaps occurred, but it comes at the expense of a doubled acquisition. In the frequency-modulated acquisitions all banding artefacts and the associated phase jumps were eliminated and no swaps between tissues occurred. It therefore provides a means to robustly separate water and fat, in one single radial bSSFP scan, using the phase-sensitive approach, even in the presence of high field inhomogeneities.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.