Abstract

This article describes the design process for a motion compensation system that can suppress the spectral distortion caused by human motion and breathing during in-vivo electron paramagnetic resonance (EPR) spectroscopy on an intact incisor. The developed system consists of two elements: an electronically controlled tunable resonator and an automatic control circuit (ACC). The resonator can modify the resonant frequency and impedance by tuning and matching the voltage, while the ACC can generate a feedback signal using phase-sensitive detection (PSD). The signal is transferred into the resonator to maintain the critical coupling state. The tunable frequency range of the resonator was measured at over 10 MHz, offering approximately eight times the required range. The bandwidth of the resonator fluctuated in a negligible range (0.14% relative standard error) following the resonant frequency. With the feedback signal on, in-vivo EPR measurements were demonstrated to be a stable baseline with 35% higher signal-to-noise ratio (SNR). When one incisor sample was irradiated by an X-ray instrument, the EPR signal responses to the absorbed doses of 0–10 Gy exhibited high linearity (R2 = 0.994). In addition, the standard error of inverse prediction was estimated to be 0.35 Gy. The developed system achieved a discrimination ability of 2 Gy, which is required for triage in large-scale radiation accidents. Moreover, the compensation is fully automated, meaning that the system can be operated with simple training in an emergency.