Abstract

The diagnostic performance of the frequency doubling perimetry (FDT) C20-1 screening mode was compared to that of the C20 full-threshold mode. For the number of defects p < 1% in the total deviation plot, both modes appeared to perform similarly in terms of sensitivity, specificity, and area under the receiver-operating characteristic (ROC) curve. Different cut-off points should be applied for both modes to obtain equal sensitivity and specificity values, and - related to that - for most subjects more defects were found in full-threshold mode than in screening mode. For the screening mode, we found a sensitivity of 0.91 and a specificity of 0.88 at a cut-off point of >0 defects, and an area under the ROC curve of 0.93.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.