Abstract

In freehand elastography, quasi-static tissue compression is applied through the ultrasound probe, and the corresponding axial strain is estimated by calculating the time shift between consecutive echo signals. This calculation typically suffers from a poor signal-to-noise ratio or from the decorrelation between consecutive echoes resulting from an erroneous axial motion impressed by the operator. This paper shows that the quality of elastograms can be improved through the integration of two distinct techniques in the strain estimation procedure. The first technique evaluates the displacement of the tissue by analyzing the phases of the echo signal spectra acquired during compression. The second technique increases the displacement estimation robustness by averaging multiple displacement estimations in a high-frame-rate imaging system, while maintaining the typical elastogram frame-rate. The experimental results, obtained with the Ultrasound Advanced Open Platform (ULA-OP) and a cyst phantom, demonstrate that each of the proposed methods can independently improve the quality of elastograms, and that further improvements are possible through their combination.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.