Abstract

This research investigates the critical role of frequency selection in Electrical Impedance Tomography (EIT), a non-invasive imaging technique that reconstructs internal conductivity distributions through injected electrical currents. Empirical frequency selection is paramount to maximizing the fidelity and specificity of EIT images. The study explores the impact of distinct frequency ranges—low, medium, and high—on image contrast and clarity, particularly focusing on differentiating conductive materials from non-conductive materials. The findings reveal distinct empirical frequency bands for enhancing the respective contrasts: 15–38 kHz for conductive materials (copper) and 45–75 kHz for non-conductive materials (acrylic resin). These insights shed light on the frequency-dependent nature of material contrast in EIT images, guiding the selection of empirical operating ranges for various target materials. This research paves the way for improved sensitivity and broader applicability of EIT in diverse areas.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.