Abstract

If dispersion in a medium is weak and approximately linear with frequency (over the experimental band of frequencies), then it can be shown that the constant term in a polynomial representation of phase shift as a function of frequency can produce errors in measurements of phase-velocity differences in through-transmission, substitution experiments. A method for suppressing the effects of the constant phase shift in the context of the single-wave-model was tested on measurements from 30 cancellous human calcaneus samples in vitro. Without adjustment for constant phase shifts, the estimated phase velocity at 500 kHz was 1516+/-6 m/s (mean+/-standard error), and the estimated dispersion was -24+/-4 m/s MHz (mean+/-standard error). With adjustment for constant phase shifts, the estimated mean velocity decreased by 4-9 m/s, and the estimated magnitude of mean dispersion decreased by 50%-100%. The average correlation coefficient between the measured attenuation coefficient and frequency was 0.997+/-0.0026 (mean+/-standard deviation), suggesting that the signal for each sample was dominated by one wave. A single-wave, linearly dispersive model conformed to measured complex transfer functions from the 30 cancellous-bone samples with an average root-mean-square error of 1.9%+/-1.0%.