Abstract

The readout noise of a H2RG HgCdTe NIR detector from Teledyne is measured at a temperature T=110K. It is shown that a Fowler mode with n=240 allows to reach a noise of 2.63e (single read). A description of the power spectrum in terms of three parameters reproduces the variation of the noise as a function the number of Fowler samples as well as its dependence on the periodicity of the sampling. The variance of the noise decreases with frequency with an effective power of 0.62 in our measurement domain. The behaviour of the detector under different experimental conditions can then be predicted.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.