Abstract
In a recent 99m Tc-HYNIC-PSMA study conducted at our department, we examined 2 patients with prostate cancer referred for initial staging on the same day. The whole-body scans revealed radiotracer uptake in the gastric mucosa and thyroid glands, alluding to high levels of free TcO 4- in the injected vial. The scans were repeated after confirming acceptable radiopharmaceutical purity of 97% (normal range, 95%-100%). Interestingly, 1 patient had liver metastases at presentation, which remained non-PSMA-avid after repeating the scan. We have reviewed this pitfall, which has been reported with many radiotracers, yet not reported with PSMA tracers.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.