Abstract
Surprisingly, a reliable method for measuring the concentration of free fluoride ions in honey is still missing from the literature, notwithstanding the generally recognized importance of the analyte and the matrix. To fill this gap, this study proposes and validates a straightforward ion-specific electrode potentiometric method for this task. The method offers very low detection and quantification limits (6.7μgkg−1 and 25μgkg−1, respectively), good linearity (R2>0.994), good sensitivity (typically 55±3mV for an order of magnitude of concentration) in an unusually low concentration interval (between 0.020 and 1mgL−1), and acceptable precision and bias. The method was applied to 30 unifloral (thistle, eucalyptus and strawberry tree) honey samples from Sardinia, Italy. The amount of free fluoride ions found in these honeys appears to be lower than the range usually found in the literature; indeed, early results suggest a possible dependence of the analyte concentration on the honey’s botanical origin.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.