Abstract

PurposeTo demonstrate free-breathing thoracic MRI with a minimal-TR balanced steady-state free precession (bSSFP) technique using wobbling Archimedean spiral pole (WASP) trajectories. MethodsPhantom and free-breathing in vivo chest imaging in healthy volunteers was performed at 1.5T with a half-radial, dual-echo, bSSFP sequence, termed bSTAR. For maximum sampling efficiency, a single analog-to-digital converter window along the full bipolar readout was used. To ensure a homogeneous coverage of the k-space over multiple breathing cycles, radial k-space sampling followed short-duration Archimedean spiral interleaves that were randomly titled by a small polar angle and rotated by a golden angle about the polar axis; depticting a wobbling Archimedean spiral pole (WASP) trajectory. In phantom and in vivo experiments, WASP trajectories were compared to spiral phyllotaxis sampling in terms of eddy currents and were used to generate in vivo thorax images at different respiratory phases. ResultsWASP trajectories provided artifact-free bSTAR imaging in both phantom and in vivo and respiratory self-gated reconstruction was successfully performed in all subjects. The amount of the acquired data allowed the reconstruction of 10 volumes at different respiratory levels with isotropic resolution of 1.77mm from a scan of 5.5minutes (using a TR of 1.32ms), and one high-resolution 1.16mm end-expiratory volume from a scan of 4.7minutes (using a TR of 1.42ms). The very short TR of bSTAR mitigated off-resonance artifacts despite the large field-of-view. ConclusionWe have demonstrated the feasibility of high-resolution free-breathing thoracic imaging with bSTAR using the wobbling Archimedean spiral pole in healthy subjects at 1.5T.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call