Abstract

The flow re-direction endoluminal device (FRED) is a safe and effective treatment option for intracranial aneurysms. The novel FRED X features an antithrombotic surface coating ("X Technology") on an otherwise unmodified stent design. This two-center study evaluates the clinical safety and efficacy of FRED X and compares it to the literature. Consecutive patients treated between 2020 and 2023 were retrospectively reviewed for aneurysm characteristics, procedural details and complications, and angiographic outcomes. A mini-review of the literature for FRED X clinical trials was performed and results were pooled using a random effects model. Thirty-four patients (mean age 56 years) were treated for 34 aneurysms. The mean aneurysm size was 7.7 ± 5.0 mm, 7 (21%) were ruptured, 6 (18%) were recurrent after previous treatment, 11 (32.3%) were located in the posterior circulation, and 4 (12.5%) had non-saccular morphology. All procedures were technically successful and no balloon angioplasty was required. There was 1 (2.9%) symptomatic complication (a transient ischemic attack) and no procedural morbidity or mortality. Technical asymptomatic events included 1 procedural stent occlusion that was reopened with thrombectomy and 3 cases of vasospasm. Complete and adequate occlusion rates were 68% (19/28) and 89% (25/28) at a mean follow-up time of 6 months, respectively. The results of this study are comparable to previous FRED X studies. The results demonstrate a high feasibility and procedural safety of the FRED X with adequate mid-term occlusion rates. Long-term and comparative studies are needed to evaluate the full potential of the FRED X.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call