Abstract

A clinical study was conducted of 59 patients treated with 78 hydroxyapatite (HA)-coated blade implants from August 1986. Five implants in 5 patients were broken at the neck portion, and one implant in one patient was removed from the jaw bone. The histological findings around a broken implant which was removed from the mandible are presented. These sections showed good adaptation of the bone to the implant without a fibrous layer. Histomorphometric evaluation of bone-to-implant contact showed 73.5%. The scanning electron microscopy image of the fractured surface revealed a fatigue fracture. The suspected cause of the fracture was stress concentration at the cervix portion, because of excessive mobility or deleterious change of abutment teeth. Consequently, the osseointegration/biointegration implants should not be connected with natural teeth.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.