Abstract
We measured healing in 7 cases of surgical neck fracture of the humerus by applying a strain gauge measuring bar between the external fixation pins. Repeated measurements provided a healing curve for the individual cases. The gradual decrease in deflection of the fracture zone reached a plateau of about 50 percent after 2 to 4 weeks in 6 cases. Removal of the frame at this time proved safe, as solid healing occurred. In one fracture no such decrease was seen and nonunion developed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.