Abstract

We present a rare case of fracture of the modular femoral component in a varus-valgus constrained total knee arthroplasty (TKA) prosthesis which occurred 2 years after implantation. Both femoral and tibial components underwent a second revision using a constrained hinged TKA. Laboratory analysis revealed insufficient metaphyseal bone stock in association with inadequate cement mantle incorporation which made implant fixation solely reliant on the intramedullary stem construct. It was therefore hypothesized that high repetitive stresses over the junction of the modular component led to the fracture. Case report, Level IV.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.